索引于
  • 打开 J 门
  • Genamics 期刊搜索
  • 学术钥匙
  • 期刊目录
  • 全球影响因子 (GIF)
  • 中国知网(CNKI)
  • 乌尔里希的期刊目录
  • 参考搜索
  • 哈姆达大学
  • 亚利桑那州EBSCO
  • OCLC-WorldCat
  • 普布隆斯
  • 日内瓦医学教育与研究基金会
  • 欧洲酒吧
  • 谷歌学术
分享此页面
期刊传单
Flyer image

抽象的

Actuator Performance Comparison by an Integrated Multivariate Approach

Copelli D, Bodria A, Magnani I, Militerno G, Ponticelli M, Usberti F and Leardi R

During the development of a new Pressurized Metered Dose Inhalers (pMDI) one of the key aspect is to achieve the right performances of the emitted aerosol. Among all variables that can impact the performances, the configuration of the actuator largely affects the atomization process and, as a consequence, the pMDI performances. In order to understand the effect of actuator orifice diameter and actuator sump volume on the final performance, a Face Centered Design has been applied. For each experiment fifteen responses were measured, using a Unit Spray Collector Apparatus (USCA) for the Delivered/Metered Dose and a Next Generation Impactor (NGI) for the Aerodynamic Particle Size Distribution tests. A Principal Component Analysis showed that the NGI responses can be used to demonstrate how the two variables affect the performance of the actuators, while the USCA responses are useless for this purpose. The effect of orifice diameter and sump volume on Fine Particle Mass, Fine Particle Fraction and Mass Median Aerodynamic Diameter, three relevant Aerodynamic Particle Size Distribution responses, were then quantitatively evaluated by Multiple Linear Regression.

免责声明: 此摘要通过人工智能工具翻译,尚未经过审核或验证