索引于
  • 打开 J 门
  • Genamics 期刊搜索
  • 学术钥匙
  • 期刊目录
  • 全球影响因子 (GIF)
  • 中国知网(CNKI)
  • 乌尔里希的期刊目录
  • 参考搜索
  • 哈姆达大学
  • 亚利桑那州EBSCO
  • OCLC-WorldCat
  • 普布隆斯
  • 日内瓦医学教育与研究基金会
  • 欧洲酒吧
  • 谷歌学术
分享此页面
期刊传单
Flyer image

抽象的

The Synergy of Electrochemical Impedance Spectroscopy and Medicinal Technology

Edward P Randviir and Craig E Banks

A reversed phase high performance liquid chromatographic assay coupled with UV detection [239 nm] has been developed and validated for the determination of midazolam in rat plasma samples after transdermal administration. A liquid–liquid extraction was used to extract the compound from plasma sample. The separation was performed on a Hypersil ODS C18 column using a mobile phase composed of acetonitrile and 0.1% triethylamine aqueous solution [ 52:48, V/V], pumped at a flow rate 1.0 mL min-1. The calibration curves showed good linearity with correlation coefficient higher than 0.998 for all analytes in the range 0.10–10.0 μg mL-1. Accuracy in the measurement of quality control [QC] samples was in the range 95–107% of the nominal values. The intra-day and inter-day precisions in the measurement of QC samples were less than 10% coefficient of variation. The developed method is suitable for quality control of midazolam in their mixtures and in transdermal delivery system pharmaceutical preparations. The validated assay was found to be suitable for the pharmacokinetic study of midazolam in rats with transdermal administration.